Flow controllers and RGA setup
Instructions
Written by Nebojsa Marinkovic (Ned)
Two identical systems are available, one for each hutch.
The procedures below explain operation on any of them.
Up to four gases can be simultaneously regulated and/or
mixed using the flow controller setup. If you need to connect more than four
gases at any time, check with beeline staff.
Mass Flow Controllers and Purification system
operation
- Go
inside the hutch and locate: a) the plate with four Brooks mass flow
controllers (MFCs) and b) gas purification rack. In the X18b hutch, both are
placed below the Io ion chamber. In the X19A hutch, the mass flow
controllers’ plate is sitting on the lower shelf next to the hutch door
(on the left hand side), and the purification system is on the left wall
(see Figs. 1 and 2).
Figure
1. (below left) Mass
flow controllers in X19A hutch, showing color-coded gas inlets. Brown stainless
steel tubing is reserved for He through purifiers, other three can vary. The MFC
outlet (white PVC tubing) should be connected to your reactor system.
Figure
2. Gas Purification
system with moisture cartridges (first two on the left) and oxygen cartridges
(two on the right).
- Check
that all MFCs are connected to tubes and all inlet valves before MFCs are
open. Remember the order of color tubes (green, red, yellow and brown in
Fig.1). The brown stainless steel tubing is used for the carrier gas and
goes through the purification system. We normally use He as the carrier gas.
Thus, the purifier cartridges are saturated with He. If you want another
carrier gas, check with beamline staff. Also, check that each MFC has a grey
electrical cable connected.
- Rotate
the input 3-way valve on the gas purification rack to ‘Bypass’ as in
Fig.2.
- Outside
the hatch, connect all gases to color tubes. Make sure that the pressure on
the gas regulator is at least 17 PSI (the check valves before MFCs open at
pressures reater than 14.7 PSI, or 1 atm).
- Check
that the electronic MFC regulator box outside the hutch (Fig. 3) has four
gray cables connected on the back. Turn the box on.
Figure 3.
Electronic MFC regulator box on X18b.
Each
channel on MFC regulator box (above) controls one MFC. Channel 4 is reserved for
the carrier gas; other three can vary by the user. Channel 1 is the one closest
to you (the first one in Fig.1, connected to the green tubing).
- Test
the flow controller electronic box. Press channel selector key until
‘ > ’ symbol appears next to Channel 4. Then, press
‘Open,close, reg.flow’ key repeatedly until it changes to ‘ V+ ’.
and quickly press ‘Enter’ key. The valve is now fully open, i.e. the
flow rate is > 50 mL/min. If there is flow through the MFC #4, it will be
displayed.
- While
channel selector key still points to Channel 4, press ‘Open,close,
reg.flow’ key repeatedly until it displays ‘ V0 ‘ and press
‘Enter’. The MFC is now in regulated mode. Using the up or down arrow
keys, select the desired flow and press ‘Enter’. The flow is expressed
as the percentage of the maximum flow rate of the MFC (50 mL/min). Thus, 20%
corresponds to 10 mL/min, 30% corresponds to 15 mL/min, etc.
- Close
Channel 4 by pressing ‘Open,close, reg.flow’ key repeatedly until it
displays ‘ V- ‘ and press ‘Enter’. Move the selector key to another
used channel (e.g. channel 1) and test its flow. Test all other used
channels using the procedure above.
- If
you want to use purification system, rotate its three-way valve to ‘On’.
Open the two slave valves by rotating them 90o so that they are
vertical, i.e. in line with the tubing (both are closed in Fig. 2). Remember
to close them when you no longer need the purifiers, or at the end of your
run.
Residual
Gas Analyzer setup and operation
Figure 4.
Residual Gas Analyzer
Residual gas analyzer (RGA) is usually on and ready for
use. In an unlikely case that you find it off, it may need several hours before
it gets in operating mode. You may want to check with the neighboring beamline
users (X18B or X19A) to see if they are using the RGA and whether it is on. If
their RGA is ready but not being used, you can roll it to your beamline and use
it within an hour. If you need to move the RGA, start from ‘turning RGA
off’. If you want to power on your RGA, start with ‘setup’. Start with
‘operation’ if it is ready.
Powering RGA off
The turbo pump must stop spinning before you can move the
RGA cart. It takes about 30 min to spin it down.
- Close
the vacuum valve by rotating it 90o clockwise.
- Turn
off the whole unit by pressing the main switch on the power strip on the
bottom of the cart. After the rotary pump stops, the only sound coming from
the system is your turbo pump. Wait until you can hear the sound no longer
(about 30 min).
- Disconnect
gas lines connected to the RGA cart. Input line is connected through the
filter just left to the flow meter, exhaust line is connected to the bubbler
(see Fig.4). Unplug the power strip and the RS232 cable from the RGA head.
- Roll
the RGA to your beamline and connect it to the input and exhaust gas lines.
Connect the RS232 cable to the RGA head.
Setup
1.
Check that the power strip on the bottom of the RGA cart is plugged in.
Turn the power strip switch on. The rotary pump starts.
2.
Wait for about one minute, than press ‘start/stop’ button on the
turbo pump controller (see Fig. 4). Turbo pump starts. The controller displays
the rotation of the turbo pump as it accelerates. Once it reaches 56,000 rpm,
the display shows ‘normal operation’. Wait another five minutes before
continuing with step 3.
3.
Open vacuum valve (black round valve, Fig 4) by rotating it 90o
anticlockwise, so that its line is vertical.
4.
On the ‘pressure readout’ box (fig. 4), press the red button. After
its initialization (about 15 sec) the readout shows the pressure next to Pirani
gauge 1 (PG 1). The Pirani gauge displays pressures from atmospheric pressure
down to 10-4 torr. If the pressure is lower than its limit, it shows
zero.
5.
If PG 1 reads zero, press the top black button on the ‘pressure
readout’ box labeled ‘IG 1’. Ion gauge is now on. After a few seconds it
displays the pressure in the RGA. Depending on how long the RGA unit was off, it
may take from ½ to 24 hours before the pressure in RGA reaches 10-8
torr. Once it does, it is ready for
use.
Operation
It is important that the RGA sees very little atmospheric
gases. Make sure that you flow carrier gas (He) for at least five minutes with
the MFC fully open (V+) before opening the RGA inlet valve. Concurrently, purge
the gas lines of all gases before inputting them to the RGA.
- Do
not open the RGA input valve until you test the unit (steps 4-5). Locate the
Off/On switch on the RGA head and turn it on. Green ‘Power’ LED light
turns on.
- Trace
the RS232 cable to see which computer it is connected to. Usually it is the
users’ computer at X18b, and main (data acquisition) computer at X19A.
Locate the RGA icon on the desktop of the computer and double-click on it.
The RGA program starts.
- Click
Utilities
– RS232 Setup – Connect. If you get no error messages, your RGA
is now ready.
- Click
on the green “Go’
button on the toolbar at the top of RGA screen. You will get the message
warning you that the RGA filament must be on to proceed. Click ‘Yes’ to
turn it on. After a few seconds RGA starts collecting data.
- In
the default (Analog) mode, the screen records intensity vs. atomic mass unit
(amu). You will get a screen with peaks at 2, 18, 28, 32 and 44 amu,
corresponding to H2, H2O, N2, O2
and CO2 gases (see Fig. 5). As the filament degases, all these
peaks diminish. Depending on how long was the filament off, it may take 5 to
30 minutes before the intensity of the peaks reach steady state. It is not
necessary to wait for the steady state, you can open the RGA input valve
using it after 5 min.

Figure
5. Typical RGA
spectrum with input valve closed. Major peaks are found at mass units of 2 (H2),
18 (H2O), 28 (N2) and 44 (CO2 and/or pump oil).
- With
He gas flowing, carefully open the RGA input valve (see Fig. 4), making sure
not to touch the glass capillary it is connected to. After the initial burst
of N2 and O2 peaks (due to air trapped in the
capillary), your spectrum should start to show only one major peak at 4 amu.
- A
powerful feature of RGA program is to view the changes in partial pressure
of gases as a function of time. Choose Mode – P
vs. T. By default, the program registers only N2, but one
can set up to 10 different gases or gas fractions by going into Scan
– Table Parameters. The program can check the partial pressure
every second (default) or it can accept user-defined time interval (Scan
– Schedule).
- It
may be a good idea to view/record P vs. T along with another scan, e.g.
Analog scan (c.f. Fig. 5). If you are still in P vs. T mode, click Mode
– Split – Analog. The window divides in two portions, the top
showing the Analog scan, the bottom P vs. T. In the split mode, the time
interval is defined by the longer of the two modes displayed, i.e. the
Analog scan; the minimum time interval to scan 50 amu (with default speed)
is 24 seconds. The speed can also be changed in Scan
– Schedule.
- Data
can be saved as a .RGA file, readable only by the RGA program, or it can be
saved as a text (ASCII) file that can be processed later with a plotting
program. To save it as a text, click File – Save As – ASC text.
If you use split screen, the data will be saved twice, first as the top
window (Analog scan), second as the bottom window (P vs. T). Make sure you
give different names (up to 8 characters) for the two scans. You do not have
to stop the scan while saving – the program will save all points it has in
memory from the time the scan was initiated.
- Fig.
6 shows P vs. T scan of a 15 hour-long run with successive oxidation (with
10% O2 in He) and reduction (with 5% H2 in He) of Cu
in CeO2, saved as an ASC file and plotted using Microsoft Excel.
Fluctuations in partial pressures of He H2 and O2 are
clearly visible, as well as the increase in water level during the reduction
process.

Figure
6. P vs. T output of
the RGA program, showing partial pressures of gases during oxidation and
reduction of Cu in CeO2, plotted using Microsoft Excel.